Repair Log: Tektronix 2445 Oscilloscope

My Tektronix 2445 Oscilloscope had been acting up for awhile: sometimes it would properly powerup when I pressed the power-on button, and sometimes it would remain off, irrespective of the ON button's state, only to turn ON after repeated toggling of the power pushbutton.

And then, finally, it stopped working -- no amount of toggling the power pushbutton would cause it to come on -- there was no trace on the CRT, the fan wasn't turning...it was dead.

Primary suspect was the power supply -- electrolytic capacitors can fail.  Could this be the problem?  Or was it something more basic?  Was the power supply getting power?

A quick check of the fuse showed it was good.  And, although electrolytic capacitors are always a suspect, it was possible that the power switch was bad (given that the oscilloscope was completely dead).

The schematic below shows the path through the Power Switch from the AC power connector to the AC input pins of the bridge rectifier CR1011.  The first thing to check would be if there existed a low-resistance path between each power pin of the AC power connector and the appropriate CR1011 AC input pin, with power switch S350 closed, per the schematic, below.


To check these paths I would have to remove the scope's cabinet, the top aluminum cover that sits over the chassis, and the plastic primary circuit shield that sits over the rear of the power-supply assembly.

The Tek 2445 manual includes instructions for disassembling the oscilloscope: refer to page 6-22 for removal of the cabinet, page 6-23 for removal of the top cover, and page 6-25 for removal of the power-supply assembly (consisting of boards A2, A3, and A4) and the plastic primary circuit shield.

Using an ohm-meter to measure the resistances between the AC power connector pins and the AC input pins of CR1011, it was obvious that lower AC path (per the schematic, above) was open, even though the switch was closed.


To determine what part in the path had failed, I removed the power-supply assembly from the chassis and then removed the A2 PCB (Regulator Board) from the assembly.

Probing this board revealed that the power switch S350 was bad.  I removed this switch and took a closer look at it by removing its top cover.  One of the contact pins had overheated -- so much so that it had melted the plastic around it and the switch pole could no longer make proper contact with the pin.  See below (upper left contact):


So the switch (ref. designator S350) needed replacing.

Per the manual, the switch was manufactured by IEE/SCHADOW INC.  Part number is NE15/F2U103EE (DPDT switch rated at 4A, 250VAC).  Some googling will get you a number of links that discuss its replacement, such as:  https://www.eevblog.com/forum/repair/schadow-ne15-(hp)-power-switch-replacement-or-alternative/

So, two alternative switches were:

  • Mouser:  WELL BUYING, PT-S4PL (Mouser P/N: 107-PT-S4PL, rated 4A, 250VAC).
  • Digikey:  C&K, NE182UEE6AMP (note that this is rated at 6 Amps)

A quick check revealed Mouser had their switch in stock but not Digikey.  So I placed an order with Mouser for the PT-S4PL.

Note that the original switch was covered with a small plastic housing -- I assume for safely reasons. (The photo below shows this housing after I removed it from covering the switch).


When the new switch arrived I soldered it onto the board and reinstalled the small plastic housing over it.  

To make sure everything was copacetic before I completely reassembled the scope, I first installed PCB A2 back into the power-supply assembly and this assembly back into the chassis.  I attached power to the scope and powered it on.  A trace appeared on the CRT and the fan was turning.  

Hurray!  It works!

Other Notes:

1. One question is why did the power-switch pin overheat?  A couple of possibilities -- one was that there was too much current, and another was that the switch contacts were not forming a low-resistance path.

A likely suspect was that one or more electrolytic capacitors in the power supply had become very lossy.  I checked these in situ (but with PCB A2 removed from the power-supply assembly and that assembly removed from the oscilloscope's chassis) using a GenRad 1657 RLC Digibridge.  All capacitors that I measured looked fine both from a capacitance and a loss perspective.

So, as of this moment, I do not know why the switch overheated.